Safety instructions
BOWA-IFU-13120-ERGOECO-S0-EN-20220420 Instructions for Use ERGOeco
9
Comply with the Instructions for use of the HF device and the general instructions for
electrosurgical operations!
•
If necessary, use a suitable tester (e.g. BOWA REF 050-230) to check if the device is in
proper working order.
•
The "BF" / "CF" applied part of the HF device used is extended by the instrument
connected to it.
3.2 Use-related
•
Make sure that the HF generator you intend to use is connected to a matching instrument
handle (see chapter 1.1 “Scope of validity”).
•
Before you use the device, run a function test using a gauze soaked in saline solution.
•
Only insert the bipolar instrument into the body if you have a clear field of vision.
•
Ensure that you have adequate visibility before activating the jaws .
•
For coagulation in the vicinity of sensitive structures, such as nerves or the ureter,
proceed with the utmost caution .
•
The jaws must not be too full. Never grasp too much tissue for coagulation.
•
Improper use of HF current can lead to burns and explosions:
Perform electrosurgical operations only with insufflation of non-flammable gases
(CO
2
).
Avoid direct skin contact with HF cables.
Avoid contact with flammable gases and liquids.
•
Improper use of the device can cause injury to the patient:
Avoid skin-to-skin contact between the HF cable and the patient.
When plugging in or removing the HF cable, always take hold of the connector, never
pull on the cable itself.
To prevent injury due to inadvertant activation of the HF,do not place the instrument
on the patient.
Avoid contact with metallic objects (clips, stents, etc.) in the vicinity of the active jaws.
This can affect energy output and lead to adverse effects.
•
The jaws can get hot enough to cause a thermal lesion.
•
The HF cable may cause interference to imagery on monitors:
Never route the HF cable alongside a camera cable.
Do not lay the HF cable in loops.
Consult the Instructions for use of the BOWA HF generators and other compatible
generators for additional information on possible interference with other devices.
•
Electrical safety is improved through the use of plastic trocars.