Reprocessing
BOWA-IFU-13120-ERGOECO-S0-EN-20220420 Instructions for Use ERGOeco
17
9 Reprocessing
9.1 General information
The following values for the permitted number of reprocessing cycles are intended as a
guide. The actual number may vary depending on the demands placed on the devices.
Manual reprocessing methods are not recommended due to their significantly lower
effectiveness.
Follow the chemical manufacturers’ statements on concentrations and exposure times.
The medical device manufacturer shall not be held liable if other types of cleaning agents
and disinfectants are used.
9.2 Device-specific information
Insulated parts must not come into contact with hard, sharp or heavy devices, as this
could damage the electrical insulation.
CAUTION
Risk of infection from water splashes and steam during manual pre-
cleaning!
Wear a face mask and protective clothing.
Adequate ventilation is recommended.
NOTICE
Abrasives and metal brushes damage the jaws!
Never clean the laparoscopy instrument using abrasives or metal
brushes.
NOTICE
Compressed air causes damage to the instrument parts!
Restrict compressed air pressure to 3 bar or less for drying the
instrument parts.