UNDERSTANDING ERROR CODES AND AUDIO TONES
The Aaron 2250 includes automatic self-diagnostics. If the diagnostics detect an error, the system displays an error
code, sounds an audible tone, and deactivates the output power.
Most error codes result from faults in accessories attached to the unit. The following table lists the error codes,
describes the errors, and recommends actions to take to resolve the errors.
All error codes are displayed in the Bipolar display.
7-4
Aaron Medical
Error Code
Description
Recommended Action
F1
Cut handpiece button may be stuck
1. Turn off, then turn on the generator. Do not press buttons or activate
accessory devices during the self-test.
2. If the error code reappears, disconnect all accessories.
Turn off, then turn on the generator again.
3. If the problem persists, replace the handpiece of footswitch
and repeat the restart.
4. If the error code reappears, record the number and call
Aaron customer service.
F2
Coag handpiece button
may be stuck
F3
Cut footswitch pedal may be stuck
F4
Coag footswitch pedal
may be stuck
F5
Bipolar footswitch pedal
may be stuck
F6
Simultaneous activation error
The unit does not allow simultaneous activation of the cut and
coagulation modes. The activation mode is “first come, first served.”
This means that whichever mode is selected first will be the function
the unit is activated to dispense. An example of this functionality
includes, when the handpiece Cut button is pressed, the unit is activated
for Cut. If a footswitch is simultaneously pressed for Coag, the unit will
continue in the Cut mode as long as the handpiece Cut button is
pressed. If the Cut button is released, the unit will sense an error
and both functions will be disabled.
1. Release either the cut or coag button on the handpiece,
or the cut or coag pedal on the footswitch.
2. If the error code reappears, record the number and contact
Aaron customer service.
E1
Output current out of specification
1. Turn the unit off.
2. Turn the unit on.
3. If the error code reappears, record the number and contact
Aaron customer service.
E2
Dosage voltage error
E3
Dosage current error
E4
DC power error
E5
Internal temperature of a section of
the unit exceeded the limit.
1. Turn the unit off.
2. Allow the unit to cool for 20 minutes.
3. Turn the unit on.
4. If the error code reappears, record the number and contact
Aaron customer service.
E6
E7
E8
NEM internal error
1. Turn the unit off.
2. Turn the unit on.
3. If the error code reappears, record the number and contact
Aaron customer service.
Summary of Contents for AARON 2250
Page 1: ...i Service Guide Aaron 2250 SERVICE GUIDE...
Page 8: ...viii Aaron Medical...
Page 14: ...1 6 Aaron Medical...
Page 54: ...5 6 Aaron Medical...
Page 58: ...6 4 Aaron Medical...
Page 68: ...7 10 Aaron Medical...
Page 73: ...B 1 Service Guide Aaron 2250 BOARD DRAWINGS SCHEMATICS ASSEMBLIES...
Page 77: ...Main Board Schematic Block Diagram B 5 Service Guide Aaron 2250...
Page 78: ...Power Factor Circuit B 6 Bovie Medical...
Page 79: ...Power Supply Circuit B 7 Service Guide Aaron 2250...
Page 80: ...SMPS Circuit B 8 Bovie Medical...
Page 81: ...Power Generator Circuit B 9 Service Guide Aaron 2250...
Page 82: ...High Voltage Relays Circuit B 10 Bovie Medical...
Page 83: ...Current Voltage Sensors Circuit B 11 Service Guide Aaron 2250...
Page 84: ...Current Voltage Sensors Backup Circuit B 12 Bovie Medical...
Page 85: ...Audio Amplifier Circuit B 13 Service Guide Aaron 2250...
Page 86: ...Display Block Diagram B 14 Bovie Medical...
Page 87: ...System Logic Circuit B 15 Service Guide Aaron 2250...
Page 88: ...Display Control Circuit B 16 Bovie Medical...
Page 89: ...Neutral Electrode Monitoring Circuit B 17 Service Guide Aaron 2250...
Page 90: ...Analog to Digital Converter Circuit B 18 Bovie Medical...
Page 91: ...Request Sense Hand Cut Circuit B 19 Service Guide Aaron 2250...
Page 92: ...Request Sense Hand Coag Circuit B 20 Bovie Medical...
Page 93: ...Request Sense Foot Cut Circuit B 21 Service Guide Aaron 2250...
Page 94: ...Request Sense Foot Coag Circuit B 22 Bovie Medical...
Page 95: ...Request Sense Foot Bipolar Circuit B 23 Service Guide Aaron 2250...
Page 96: ...Foot Monopolar Contact Circuit B 24 Bovie Medical...
Page 97: ...Foot Bipolar Contact Circuit B 25 Service Guide Aaron 2250...
Page 98: ...Silkscreen for Main PCB for Component Locations B 26 Bovie Medical...
Page 99: ...Silkscreen for Top of Display PCB for Component Locations B 27 Service Guide Aaron 2250...
Page 100: ...Silkscreen for Bottom of Display PCB for Component Locations B 28 Bovie Medical...
Page 101: ...Silkscreen for Relay PCB for Component Locations B 29 Service Guide Aaron 2250...
Page 102: ...Final Assembly Overview B 30 Bovie Medical...
Page 103: ...Cabling in Assembly Refer to page B 4 for list of cables B 31 Service Guide Aaron 2250...
Page 104: ...Mechanical Front Panel Assembly B 32 Bovie Medical...
Page 105: ...Mechanical Back Panel Assembly B 33 Service Guide Aaron 2250...
Page 106: ...Cabling Back Panel Assembly B 34 Bovie Medical...