4-8
Aaron Medical
NEM_CLK_GEN
This is and output signal from the system logic that generates a 62.5 kHz square wave.
This signal is used by the NEM circuit.
NEM_PAD_SENSE_EN
This is an output signal from the system logic that controls an analog switched. This
signal is used by the NEM circuit to change the analog signal sent to the NEM A/D.
NEM_TEMP_SENSE_EN
This is an output signal from the system logic that controls an analog switched. This
signal is used by the NEM circuit to change the analog signal sent to the NEM A/D.
SENS_RANGE
Reserved
AUD_DRV
This is an output signal from the system logic that generates the activation tones for all
modes of operation. A 1kHz square wave is generated whenever Cut I, Cut II or Blend
modes are activated. A 2Khz square wave is generated when the Coagulation,
Fulguration or Bipolar modes are activated. This signal is used by the audio circuit.
ALARM_DRV
This is and output signal from the system logic that generates a 2 kHz / 1kHz square
wave for activating the alarm siren. This signal is used by the audio circuit.
AUX_RLY
This is an output signal from the system logic that controls the accessory relay on the
back panel.
FULG_EN
This is an output signal from the system logic that controls the relays to allow fulguration
coagulation output.
MONO_EN
This is an output signal from the system logic that controls the relays to allow output to
be present at the monopolar foot controlled receptacle.
HANDLE_EN
This is an output signal from the system logic that controls the relays to allow output to
be present at the monopolar handpiece receptacle.
BIPOLAR_EN
This is an output signal from the system logic that controls the relays to allow output to
be present at the bipolar receptacle.
FOOTC_EN
This is an output signal from the system logic that controls the relays to allow output to
be present at the foot control receptacle.
RETPAD_EN
This is an output signal from the system logic that controls the relays to allow output to
be present at the monopolar return receptacle.
PFC_ON
This is an output signal from the system logic that turns on the power factor correction
circuit during activation of the unit.
Summary of Contents for AARON 2250
Page 1: ...i Service Guide Aaron 2250 SERVICE GUIDE...
Page 8: ...viii Aaron Medical...
Page 14: ...1 6 Aaron Medical...
Page 54: ...5 6 Aaron Medical...
Page 58: ...6 4 Aaron Medical...
Page 68: ...7 10 Aaron Medical...
Page 73: ...B 1 Service Guide Aaron 2250 BOARD DRAWINGS SCHEMATICS ASSEMBLIES...
Page 77: ...Main Board Schematic Block Diagram B 5 Service Guide Aaron 2250...
Page 78: ...Power Factor Circuit B 6 Bovie Medical...
Page 79: ...Power Supply Circuit B 7 Service Guide Aaron 2250...
Page 80: ...SMPS Circuit B 8 Bovie Medical...
Page 81: ...Power Generator Circuit B 9 Service Guide Aaron 2250...
Page 82: ...High Voltage Relays Circuit B 10 Bovie Medical...
Page 83: ...Current Voltage Sensors Circuit B 11 Service Guide Aaron 2250...
Page 84: ...Current Voltage Sensors Backup Circuit B 12 Bovie Medical...
Page 85: ...Audio Amplifier Circuit B 13 Service Guide Aaron 2250...
Page 86: ...Display Block Diagram B 14 Bovie Medical...
Page 87: ...System Logic Circuit B 15 Service Guide Aaron 2250...
Page 88: ...Display Control Circuit B 16 Bovie Medical...
Page 89: ...Neutral Electrode Monitoring Circuit B 17 Service Guide Aaron 2250...
Page 90: ...Analog to Digital Converter Circuit B 18 Bovie Medical...
Page 91: ...Request Sense Hand Cut Circuit B 19 Service Guide Aaron 2250...
Page 92: ...Request Sense Hand Coag Circuit B 20 Bovie Medical...
Page 93: ...Request Sense Foot Cut Circuit B 21 Service Guide Aaron 2250...
Page 94: ...Request Sense Foot Coag Circuit B 22 Bovie Medical...
Page 95: ...Request Sense Foot Bipolar Circuit B 23 Service Guide Aaron 2250...
Page 96: ...Foot Monopolar Contact Circuit B 24 Bovie Medical...
Page 97: ...Foot Bipolar Contact Circuit B 25 Service Guide Aaron 2250...
Page 98: ...Silkscreen for Main PCB for Component Locations B 26 Bovie Medical...
Page 99: ...Silkscreen for Top of Display PCB for Component Locations B 27 Service Guide Aaron 2250...
Page 100: ...Silkscreen for Bottom of Display PCB for Component Locations B 28 Bovie Medical...
Page 101: ...Silkscreen for Relay PCB for Component Locations B 29 Service Guide Aaron 2250...
Page 102: ...Final Assembly Overview B 30 Bovie Medical...
Page 103: ...Cabling in Assembly Refer to page B 4 for list of cables B 31 Service Guide Aaron 2250...
Page 104: ...Mechanical Front Panel Assembly B 32 Bovie Medical...
Page 105: ...Mechanical Back Panel Assembly B 33 Service Guide Aaron 2250...
Page 106: ...Cabling Back Panel Assembly B 34 Bovie Medical...