CONTROLS AND INDICATORS OVERVIEW
Users may control most Aaron 2250 functions from the front panel. Each control is plainly marked and
colored on the front panel for quick reference. The volume control is located on the rear panel.
Normal operations involve activating the generator with either a front connected handswitch or footswitch.
The following components are the User Interface for the Aaron 2250.
Power Switch
The rocker ON/OFF switch is located on the lower left
corner allows the 2250 to be shut off when the unit is not
in use.
Watts Displays (Cut, Coag, and Bipolar)
These large power output displays report the generator’s
output power settings from 1 to 200 watts in the Cut
mode, 1 to 120 watts in the Coag mode, and 1 to 80 watts
in the Bipolar mode all in one, five, or ten watt incre-
ments (at the rated load). During operation, the displayed
number corresponds to the selected mode, indicating the
available generator power to the surgeon.
Membrane Power, Presets,
The front panel overlay contains 10 membrane switches
& Function Switches
(sometimes called matrix switches). There is a membrane
switch dedicated to each operational mode. These switches
toggle the unit between functions (Cut, Blend, Coagula-
tion, and Fulguration) and power settings (Cut, Coag, and
Bipolar) in watts and blend settings indicated by a maxi-
mum to minimum illuminated bar).
Visual LED Indicators
Mode LEDs illuminate for Cut, Coag, Bipolar, Cut ,
Blend, Coagulation, and Fulguration settings on the front
of the unit.
The footswitch LEDs illuminates to indicate that the
monopolar or bipolar footswitch is connected to the unit.
The Power LED indicates if the unit is switched On.
The red Alarm LED indicates that the unit has detected
an error or no return electrode is attached to the
generator.
The green Return Electrode LEDs indicate the type of
return electrode sensed by the Bovie NEM™.
Audible Indicators
An activation tone sounds whenever the 2250 is activated.
The volume may be adjusted up or down on the rear of
the unit.
An Alarm Siren sounds during all alarm conditions. The
volume of this alarm cannot be adjusted.
2-3
Ser vice Guide • Aaron 2250
Summary of Contents for AARON 2250
Page 1: ...i Service Guide Aaron 2250 SERVICE GUIDE...
Page 8: ...viii Aaron Medical...
Page 14: ...1 6 Aaron Medical...
Page 54: ...5 6 Aaron Medical...
Page 58: ...6 4 Aaron Medical...
Page 68: ...7 10 Aaron Medical...
Page 73: ...B 1 Service Guide Aaron 2250 BOARD DRAWINGS SCHEMATICS ASSEMBLIES...
Page 77: ...Main Board Schematic Block Diagram B 5 Service Guide Aaron 2250...
Page 78: ...Power Factor Circuit B 6 Bovie Medical...
Page 79: ...Power Supply Circuit B 7 Service Guide Aaron 2250...
Page 80: ...SMPS Circuit B 8 Bovie Medical...
Page 81: ...Power Generator Circuit B 9 Service Guide Aaron 2250...
Page 82: ...High Voltage Relays Circuit B 10 Bovie Medical...
Page 83: ...Current Voltage Sensors Circuit B 11 Service Guide Aaron 2250...
Page 84: ...Current Voltage Sensors Backup Circuit B 12 Bovie Medical...
Page 85: ...Audio Amplifier Circuit B 13 Service Guide Aaron 2250...
Page 86: ...Display Block Diagram B 14 Bovie Medical...
Page 87: ...System Logic Circuit B 15 Service Guide Aaron 2250...
Page 88: ...Display Control Circuit B 16 Bovie Medical...
Page 89: ...Neutral Electrode Monitoring Circuit B 17 Service Guide Aaron 2250...
Page 90: ...Analog to Digital Converter Circuit B 18 Bovie Medical...
Page 91: ...Request Sense Hand Cut Circuit B 19 Service Guide Aaron 2250...
Page 92: ...Request Sense Hand Coag Circuit B 20 Bovie Medical...
Page 93: ...Request Sense Foot Cut Circuit B 21 Service Guide Aaron 2250...
Page 94: ...Request Sense Foot Coag Circuit B 22 Bovie Medical...
Page 95: ...Request Sense Foot Bipolar Circuit B 23 Service Guide Aaron 2250...
Page 96: ...Foot Monopolar Contact Circuit B 24 Bovie Medical...
Page 97: ...Foot Bipolar Contact Circuit B 25 Service Guide Aaron 2250...
Page 98: ...Silkscreen for Main PCB for Component Locations B 26 Bovie Medical...
Page 99: ...Silkscreen for Top of Display PCB for Component Locations B 27 Service Guide Aaron 2250...
Page 100: ...Silkscreen for Bottom of Display PCB for Component Locations B 28 Bovie Medical...
Page 101: ...Silkscreen for Relay PCB for Component Locations B 29 Service Guide Aaron 2250...
Page 102: ...Final Assembly Overview B 30 Bovie Medical...
Page 103: ...Cabling in Assembly Refer to page B 4 for list of cables B 31 Service Guide Aaron 2250...
Page 104: ...Mechanical Front Panel Assembly B 32 Bovie Medical...
Page 105: ...Mechanical Back Panel Assembly B 33 Service Guide Aaron 2250...
Page 106: ...Cabling Back Panel Assembly B 34 Bovie Medical...