4-6
Aaron Medical
RQEST_SENSE_FOOT_
BIPOLAR
This is an input signal from the bipolar footswitch sense circuit. This signal is generated
by a colpitts oscillator located on the display board and an isolation transformer located on
the main board. When an activation request is made, the output for this circuit becomes a
logic 1 (3.3 volts) signal.
RQEST_SENSE_FOOT_COAG
This is an input signal from the coag monopolar footswitch sense circuit. This signal is
generated by a colpitts oscillator located on the display board and an isolation transformer
located on the main board. When an activation request is made via the coag pedal on the
monopolar footswitch, the output for this circuit becomes a logic 1 (3.3 volts) signal.
RQEST_SENSE_FOOT_CUT
This is an input signal from the cut monopolar footswitch sense circuit. This signal is
generated by a colpitts oscillator located on the display board and an isolation transformer
located on the main board. When an activation request is made via the cut pedal on the
monopolar footswitch, the output for this circuit becomes a logic 1 (3.3 volts) signal.
RQEST_SENSE_HAND_COAG
This is an input signal from the coag monopolar hand sense circuit. This signal is
generated by a colpitts oscillator located on the display board and an isolation transformer
located on the main board. When an activation request is made via the coag button on
the monopolar handpiece, the output for this circuit becomes a logic 1 (3.3 volts) signal.
RQEST_SENSE_HAND_CUT
This is an input signal from the cut monopolar hand sense circuit. This signal is
generated by a colpitts oscillator located on the display board and an isolation transformer
located on the main board. When an activation request is made via the cut button on the
monopolar handpiece, the output for this circuit becomes a logic 1 (3.3 volts) signal.
FOOT_BIPOLAR _CONTACT
This is an input signal from the bipolar foot contact sense circuit. This signal is generated
by a colpitts oscillator located on the display board and an isolation transformer located on
the main board. When a bipolar footswitch is connected to the rear of the unit, the output
for this circuit becomes a logic 1 (3.3 volts) signal.
FOOT_MONO_CONTACT
This is an input signal from the monopolar foot contact sense circuit. This signal is
generated by a colpitts oscillator located on the display board and an isolation transformer
located on the main board. When a monopolar footswitch is connected to the rear of the
unit, the output for this circuit becomes a logic 1 (3.3 volts) signal.
NEM_DAT0 – NEM_DATA7
Data lines between the system logic and the NEM circuit. Used to monitor the
temperature of the NEM circuit and the return electrode contact impedance.
NEM_A_D_INTR
This is a communication signal from the system logic to the NEM circuit.
NEM_A_D_OFL
This is a communication signal from the system logic to the NEM circuit.
NEM_A_D_CS
This is a communication signal from the system logic to the NEM circuit.
NEM_A_D_READ
This is a communication signal from the system logic to the NEM circuit.
NEM_A_D_WRITE
This is a communication signal from the system logic to the NEM circuit.
NEM_A_D_MODE
This is a communication signal from the system logic to the NEM circuit.
Summary of Contents for AARON 2250
Page 1: ...i Service Guide Aaron 2250 SERVICE GUIDE...
Page 8: ...viii Aaron Medical...
Page 14: ...1 6 Aaron Medical...
Page 54: ...5 6 Aaron Medical...
Page 58: ...6 4 Aaron Medical...
Page 68: ...7 10 Aaron Medical...
Page 73: ...B 1 Service Guide Aaron 2250 BOARD DRAWINGS SCHEMATICS ASSEMBLIES...
Page 77: ...Main Board Schematic Block Diagram B 5 Service Guide Aaron 2250...
Page 78: ...Power Factor Circuit B 6 Bovie Medical...
Page 79: ...Power Supply Circuit B 7 Service Guide Aaron 2250...
Page 80: ...SMPS Circuit B 8 Bovie Medical...
Page 81: ...Power Generator Circuit B 9 Service Guide Aaron 2250...
Page 82: ...High Voltage Relays Circuit B 10 Bovie Medical...
Page 83: ...Current Voltage Sensors Circuit B 11 Service Guide Aaron 2250...
Page 84: ...Current Voltage Sensors Backup Circuit B 12 Bovie Medical...
Page 85: ...Audio Amplifier Circuit B 13 Service Guide Aaron 2250...
Page 86: ...Display Block Diagram B 14 Bovie Medical...
Page 87: ...System Logic Circuit B 15 Service Guide Aaron 2250...
Page 88: ...Display Control Circuit B 16 Bovie Medical...
Page 89: ...Neutral Electrode Monitoring Circuit B 17 Service Guide Aaron 2250...
Page 90: ...Analog to Digital Converter Circuit B 18 Bovie Medical...
Page 91: ...Request Sense Hand Cut Circuit B 19 Service Guide Aaron 2250...
Page 92: ...Request Sense Hand Coag Circuit B 20 Bovie Medical...
Page 93: ...Request Sense Foot Cut Circuit B 21 Service Guide Aaron 2250...
Page 94: ...Request Sense Foot Coag Circuit B 22 Bovie Medical...
Page 95: ...Request Sense Foot Bipolar Circuit B 23 Service Guide Aaron 2250...
Page 96: ...Foot Monopolar Contact Circuit B 24 Bovie Medical...
Page 97: ...Foot Bipolar Contact Circuit B 25 Service Guide Aaron 2250...
Page 98: ...Silkscreen for Main PCB for Component Locations B 26 Bovie Medical...
Page 99: ...Silkscreen for Top of Display PCB for Component Locations B 27 Service Guide Aaron 2250...
Page 100: ...Silkscreen for Bottom of Display PCB for Component Locations B 28 Bovie Medical...
Page 101: ...Silkscreen for Relay PCB for Component Locations B 29 Service Guide Aaron 2250...
Page 102: ...Final Assembly Overview B 30 Bovie Medical...
Page 103: ...Cabling in Assembly Refer to page B 4 for list of cables B 31 Service Guide Aaron 2250...
Page 104: ...Mechanical Front Panel Assembly B 32 Bovie Medical...
Page 105: ...Mechanical Back Panel Assembly B 33 Service Guide Aaron 2250...
Page 106: ...Cabling Back Panel Assembly B 34 Bovie Medical...