Configuring the CP and Stimulator
Bionic Navigator™ 3D Programming Manual
90930923-31 REV A 33 of 309
en
Measuring Impedances
Impedance values can be measured at each contact. From the Configuration - Leads screen, select
the
Impedances
button to measure impedances and display the Configuration - Impedances
screen.
Figure 21:
Impedances Screen
The results of the Impedance measurement at each contact are displayed in terms of the stimulator
port connections as seen in Figure 21. If desired, you can select the Measure button again to perform
another Impedance measurement.
A green circle indicates that the impedance measurement is less than 8000 ohms. An orange circle
indicates that the measurement is above 8000 ohms with reduced accuracy. In Figure 21, Contacts C1,