PMX40 RF Power Meter – INSTRUCTION MANUAL
4-5
Stat. Gating:
Selects Freerun or Marker gating for statistical acquisition. If
Freerun is selected, then all the samples are acquired without regard to
sweep acquisition. If Markers are selected, then only samples within the time
marker interval on the Pulse mode triggered sweep will be included in the
statistical sample population.
Term Count:
Sets the terminal sample count for the CCDF acquisition.
Term Time:
Sets the terminal running time for the CCDF acquisition.
Term Action:
Select the action to take when either the terminal count is
reached or terminal time has elapsed. Selecting Decimate will divide all
sample bins by 2 and continue. The total sample count will be halved each
time a decimation occurs. Selecting Restart clears the statistical sample population and starts a new
one. Selecting Stop will stop accumulating samples and hold the result.
4.4.5
MAIN
Channel
Channel#:
Opens the settings menu for the selected channel.
Note
The base PMX40 model only permits two sensors to be active
at any one time. With the PMX40-4CH option, four sensors
can be active at any one time.
MAIN
Channel
Channel#
Chan. Enabled:
Toggles display of the trace and measurements for the
channel. The channel may still be used as a trigger source when set to Off.
Vert Scale:
Set the power or voltage vertical axis level for the trace
display based on the units as shown in Table 4-1.
Units Scale
dBm
0.1, 0.2, 0.5 1, 2, 5, 10, 20, 50 dB/div
Watts 1 pW to 500 MW/div in a 1-2-5 progression
Volts
1 µV to 100 kV/ div in a 1-2-5 progression
Table 4-1. Vertical Scale Range for Units Selected
Vert Center:
Set the power or voltage level of the horizontal centerline of the graph for the specified
channel in the selected channel units.