PMX40 RF Power Meter – INSTRUCTION MANUAL
5-12
Table 5-3. Statistical Automatic Measurements
In the Statistical mode the following five automatic measurements are displayed in the PMX40 Text
display for both input channels and both trigger channels. The Text column lists the abbreviated forms
that appear on the display screen.
Text
TERM
DEFINITION
Avg
Average Power
The unweighted average of all linear power samples occurring since
acquisition started.
Peak
Peak Power
The highest power sample occurring since acquisition was started.
Min
Minimum Power The lowest power sample occurring since acquisition was started. In
logarithmic units a reading below the clip level will display as down
arrows.
Pk2Avg
Pk/Avg Ratio
The ratio (in dB) of the Peak Power to the Average Power.
The following six cursor measurements display the set position (independent variable) and measured
value (dependent variables) where the movable cursor intersects the measurement trace. The position
or value measurement text for each dependent variable is displayed in the color of its channel. The
independent variable is white. Note that the intersection of the movable cursors and the CCDF traces
can be moved outside the visible display area. This does not affect the measurements in any way.
Text
TERM
DEFINITION
Cursor Pwr Cursor Power
Cursor Mode - Power Ref
Reference
The reference power level in dBr set by the user to define the
measurement point on the normalized CCDF for probability in percent.
Cursor Mode - Percent
The measured power level in dBr of the normalized CCDF at the
Probability in percent specified by the user.
Cursor Pct Cursor Percentage Cursor Mode - Power Ref
The measured probability in percent of the normalized CCDF at the
reference power level specified by the user.
Cursor Mode - Percent
The probability in percent set by the user to define the measurement
point on the normalized CCDF for power level in dBr.