BOECO TC-PRO Thermal Cycler Chapter Seven Failure Analysis and Troubleshooting
—
66
—
Chapter Seven Failure Analysis and Troubleshooting
In this chapter, we briefly represent possible failures, reason analysis and troubleshooting
1. Failure Analysis and Troubleshooting
No.
Phenomenon
Failure Analysis
Troubleshooting
Disconnected
power supply
Check power supply for
correct connection
Bad fuse
Replace
it
(250V
8A
Φ
5
×
20mm)
Switch failure
Replace it
1
No display and the two beeps missed
when power on
Others
2
After power on, the instrument beeps
twice, and then the LCD display:
“
Please insert Block!
”, with beep
warning sound.
Bad connection
between
the
block and the
main body.
Power off, pull out the
block, and then re-insert
the block, and power on.
Power off, pull out the
block, and then re-insert the
block, and power on.
3
Error Message when running:
“Please
check error”
with any information
below:
“Temperature sensor disconnect”
“Temperature sensor1 disconnect ”
“Temperature sensor2 disconnect ”
“Temperature sensor3 disconnect ”
Bad
sensor
connection
or
sensor broken
Contact
distributor
or
manufacturer
Power off, pull out the
block, and then re-insert the
block, and power on.
4
Error Message when running: “
Please
check error
” with the information like
“Heat sink sensor disconnect”
Bad
sensor
connection
or
sensor broken
Contact
distributor
or
manufacturer
Power off, pull out the
block, and then re-insert the
block, and power on.
5
Error Message when running: “
Please
check error
” with the information like
“Hotlid sensor disconnect”
Bad
sensor
connection
or
sensor broken
Contact
distributor
or
manufacturer
Ventilation jam
Clear the jam
Bad connection
Open the device, check the
connectors for reliable
connection
6
Abrupt change in heating speed or
bad temperature control accuracy
Cooling module
failure
Contact distributor or
manufacturer