EN
43
4.2
Trapeze bar with triangle grip
The trapeze bar accessory weighs 6.5 kg.
The trapeze bar’s maximum safe operating weight is 75 kg.
Includes:
Trapeze bar with mounting loop
incl. triangle grip
To install, insert the trapeze bar into the socket pro-
vided on the head section of the bed and lock it into po-
sition. Hang the triangle grip through the mounting
loop.
Be sure to use only mattresses in the thickness range
recommended by Hermann Bock GmbH.
NOTE: Do not swing the trapeze bar outside of the bed
perimeter!
Under normal use conditions, the triangle grip has a
lifespan of approximately 5 years. If a trapeze bar with
triangle grip is installed on the nursing care bed, the bar
must be checked as part of each inspection and re-
placed after no more than 5 years.
The grip height can be continuously adjusted through a
range of about 350 mm. Depending on the mattress
thickness, the triangle grip can therefore be adjusted to
a height of 550 to 850 mm above the mattress surface.
When using a trapeze bar, the overall height of the
nursing care bed is increased by 1300 mm.