EN
23
3.8
Emergency lowering – Back rest (manual)
In case of a power outage or drive system
failure, you can lower the back rest by hand.
Always perform these steps with two peo-
ple!
One person lifts the back rest slightly and
holds it in the raised position. The second
person then removes the motor cover.
The motor is now separated from the back
rest and can be pushed downward. Once the
second person has left the danger zone, the
first person can carefully lower the back rest.
Be sure to keep a firm grip on the back rest
until it has been fully lowered. Before the
motor can be reassembled, it must be fully
put back in place (lower the back rest).
Bock Safety Note
This emergency lowering procedure must only be used in a true emergency, and only by
people who are able to carry it out safely. The bed must always be unplugged from mains
power until the motor has been put back in place.