EN
18
3.2
Caution: Electric drive system
With its wide variety of functions, the electrically powered nursing care bed provides the
patient with substantial physical and emotional support for their recovery process, while
also alleviating pain. As medical devices, electrically powered beds require a high level of
diligence with regard to regular safety inspections. This includes safe use of the nursing
care bed, daily inspection of electrical equipment, and proper cleaning and mainte-
nance.
To avoid cable damage, cords and cables should be placed outside of any area where dam-
age could occur. Any contact with sharp-edged components should also be prevented. To
avoid risk of injury from electric shock, any possibility of excessive touch voltage should
be eliminated. Such circumstances are especially likely to occur when the mains power
cable has been damaged, when dangerous and excessive leakage currents are present, or
when liquid has entered the motor housing, e.g. as a result of improper cleaning. Such
damage can result in control unit malfunction which can in turn lead to unintended move-
ments of individual bed components, raising the risk of injury to patients and users.
3.3
Drive systems
Hermann Bock GmbH equips its beds with drive systems from Limoss and from Dew-
ertOkin GmbH (drive system with external switch-mode power supply).
Bock Safety Note
Do not open any components of the drive system!
Only specifically authorised professionals are permitted to repair any defects or replace indi-
vidual electrical components.
Bock Safety Note
The motors have IPX4 protection against splashing water. Cables must not be pinched or
crushed. Repositioning of moveable parts may only be used for the intended purpose. Her-
mann Bock GmbH assumes no liability for unauthorised technical modifications.
Bock Safety Note
Never attempt to repair electrical equipment yourself. In some cases there may a risk of life-
threatening electric shock! Contact either the Customer Service department at Hermann Bock
GmbH or an authorised professional electrician to repair any defects in accordance with all ap-
plicable VDE guidelines and safety requirements.