10
8.3.5. Memory Card
Selecting Memory Card you go to the menu of memory.
8.3.5.1. Create backup copies of calibrations.
After calibration of the probe is recommended to create a
backup copy of the calibration (usually the manufacturer
makes the calibration of 1-2 scales, to check the probe).
This is done in order to be able to resume adequate
calibration after incorrect settings in the future.
8.3.5.2. Load backup copies of calibrations.
After the initial save of calibrations you can always
download it to the probe. This function is needed for the
resumption of adequate calibration in case of wrong
settings of the probe.
8.3.5.3. Clear SD card
Clearing saved records in the archive, after clearing SD
card the archive will be empty.
8.3.5.4. Memory state
This menu displays the total, used and empty memory
cells.
8.3.6. Information
In this menu you can view information about the
manufacturer and offices around the world. Scroll through
the list using the navigation keys
and
.
9. Hardness measurement
9.1. Using of the UCI method
The design of UCI Probe is shown in Figure 1a. Probe has a special removable
collapsible nozzle with the puck (Figure 1b). The puck provides the convenience
of positioning the probe relative to the testing object and a clip for the
measurements