G4 Multi Channel Weighing Instrument
6-45
Example 3: Set read window to read from Scale 4 floating point data
(registers 44074 to 44085). Note that the number of registers read is always 12.
1. Make sure that the previous command was not 250.
Set command byte (00) to 0 if previous command was 250.
2. Set start address (44074) in bytes 02 and 03.
See chapter ‘Communication – Modbus protocol – General registers’.
3. Enter the command (250) in byte 00.
Byte
Description
Dec
Hex
00
Command
250
FA
01
Number of registers to write
02 – 03
Start address
44074
AC 2A
04 – 15
Write registers 1 – 6.
Example 4: Reset the ‘Program reset’ bit in ‘Instrument status’.
1. Make sure that the previous command was not 252.
2. Enter the command (252) in byte 00.
Byte
Description
Dec
Hex
00
Command
252
FC
01
Number of registers to write
02 – 03
Start address
04 – 15
Write registers 1 – 6.
Example 5: Other commands like Taring, Zeroing etc.
See chapter ‘Communication – Modbus protocol – Command register’ for
a list of available commands.
1. Make sure that the previous command was not the desired command.
2. Enter the desired command in byte 00.
Byte
Description
Dec
Hex
00
Command
XXX
XX
01
Number of registers to write
02 – 03
Start address
04 – 15
Write registers 1 – 6.
Note that a dash in Dec. or Hex columns above indicates that the value is unimportant
and not used for the function described.
Summary of Contents for G4-DT
Page 1: ...G4 Multi Channel Weighing Instrument Program version 1 12 0 0 Technical Manual PM DT HE types ...
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 36: ...Technical Manual 2 18 ...
Page 41: ...G4 Multi Channel Weighing Instrument 3 5 ...
Page 154: ...Technical Manual 6 48 ...
Page 180: ......
Page 190: ...Technical Manual 9 10 ...
Page 191: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 192: ...Technical Manual ...
Page 193: ......