G4 Multi Channel Weighing Instrument
6-35
The first 32 bytes of data from the instrument to the network are mandatory.
Bytes 0 and 1:
Contains the actual instrument error information (the value 0 means no error).
See chapter ‘Communication – Modbus protocol – Instrument error’.
Bytes 2 and 3:
The actual status information for the instrument,
see chapter ‘Communication – Modbus protocol – Instrument status’.
Byte 4:
The ‘Command acknowledge’. It will be equal to the command number if
the command was successfully executed. If the command failed byte 4 will have
the value 240 (hex F0).
Byte 5:
The possible ‘Command error’ code. The error code is explained in chapter
‘Troubleshooting – Error codes’. This byte will be zero if a command is correctly
executed.
Bytes 6 and 7:
These two bytes is the starting address of the reading area following in bytes 8 to 31.
The address defines the first of the 12 consecutive modbus registers in the read area.
See chapter ‘Communication – Modbus protocol’ for details on modbus register
address definitions.
Bytes 8 to 31:
These 24 bytes is the area used when the user wants to read any data that is
accessible through modbus registers. The address (number) of the first register is
defined in bytes 6 and 7 and the instrument will keep the 24 bytes (12 registers)
updated.
Summary of Contents for G4-DT
Page 1: ...G4 Multi Channel Weighing Instrument Program version 1 12 0 0 Technical Manual PM DT HE types ...
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 36: ...Technical Manual 2 18 ...
Page 41: ...G4 Multi Channel Weighing Instrument 3 5 ...
Page 154: ...Technical Manual 6 48 ...
Page 180: ......
Page 190: ...Technical Manual 9 10 ...
Page 191: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 192: ...Technical Manual ...
Page 193: ......