Fitting Advice
9.1
Moisture in the Distal End of the Device
If moisture collects inside the distal end of the device, refer to the following table:
Cause
Solution
The distal perforations are clogged.
Clean the device. (See
Maintenance
section.)
The socket does not have enough space in its
distal end for the valve to open.
Put the spacer between the pin lock and the
valve. (See
Fitting the Spacer
section.)
If unsuccessful cast another socket.
The internal shape of the lock-body is
stopping the valve from opening.
Put the spacer between the pin lock and the
valve. (See
Fitting the Spacer
section.)
9.2
Loss of Vacuum
If the device loses vacuum, refer to the following table:
Cause
Solution
The valve is damaged.
Stop using the device, and contact a
Blatchford sales representative.
Debris beneath the valve preventing correct
seal.
Clean the valve. (See
Cleaning the Valve
section.)
The locking pin is not aligned with the long
axis of the residual limb.
Re-align the locking pin
with the long axis of
the residual limb.
9.3
Separation of the Valve from the Distal Cap
If the valve falls off the distal cap, refer to the following table:
Cause
Solution
The valve is not fitted correctly.
Make sure that the washer is between the pin
and the valve. (See
Fitting the Valve
section.)
Note… Do not over-tighten the locing pin/lock rod.