Figure 2.8
System Config Menu
Figure 2.9
System Self Test
2.8.1 Setup Info Menu
From the top level after pressing the “Menu” key, the second soft-key is “Setup”. This menu contains settings for
configuring the measurement system of the meter. Setting such as averaging, external sensors, update speed and similar
are configured under this menu. See
2.8.2 Averaging Setup
When measuring low frequency signals where measurements begin to become unstable, averaging may be useful. To
enable averaging enter the “Average Setup” menu (Menu>Setup>Average Set). See
.
2.8.2.1 EXP - Index averaging
𝐷
𝑛
= 𝐷
𝑛−1
+
𝑀
𝑛
− 𝐷
𝑛−1
𝑘
Dn
The value displayed after the nth index averaging (D1, the value displayed after the first aver-
aging, is equal to M1)