124
If Quick-Cal has been turned On previously and the current vertical scale is 2 mV/div or 5 mV/div
on any channel, the oscilloscope will perform the quick calibration procedure when it powers on.
During the calibration, most of the buttons are disabled.
When Quick-Cal is On and the current vertical scale is 2 mV/div or 5 mV/div on any channel, the
oscilloscope will perform the quick calibration procedure if the operating temperature changes
by more than 2 °C.
11.8
Update Firmware and Configuration
The firmware and configuration can be updated via the USB storage device. Follow these steps
to update the firmware:
1.
Insert the USB flash drive containing the firmware file into the USB host connector on the
front panel.
2.
Press the Utility button to enter the UTILITY menu.
3.
Press the Next Page softkey twice to go to the third page of the UTILITY function menu.
4.
Press the Update softkey to enter the UPDATE menu.
5.
Press the Firmware softkey and you will see the message “Press ‘Single’ to continue and press
‘Run/Stop’ to exit!”.
6.
Press the Single button to enter the SAVE/RECALL file system.
7.
Turn the Universal Knob to select the update file (it should have an ADS suffix). Press the
Load softkey to start loading the new firmware. The process takes about 7 minutes.
IMPORTANT: Do not turn off the oscilloscope during this process or the oscilloscope will not
start again.
8.
After finishing the update, the screen will display the message “Update success, please restart
the DSO”. The oscilloscope will make a buzzing sound.
9.
Turn the power off and back on again to finish the firmware update.
Follow these steps to update the configuration:
1.
Insert the USB flash drive containing the configuration file into the USB host interface on the
front panel of the oscilloscope.
2.
Press the Utility button to enter the UTILITY menu.
3.
Press the Next Page softkey twice to go to the third page of the UTILITY function menu.
4.
Press the Update softkey to enter the UPDATE menu.
5.
Press the Configure softkey to enter the SAVE/RECALL file system.
6.
Turn the Universal Knob to select the configuration file which should have a CFG suffix and
press the Load softkey to start update the firmware. The process takes about 30 seconds.
7.
After finishing the update, the screen will display the message “Update success, please restart
the DSO”. The oscilloscope will make a buzzing sound.
8.
Turn the power off and back on again to finish the configuration update.