Menu Operation
83
Press
UTILITY
→
Self-Cal
to display the self-calibration
page. Press
AUTO
key to exit the Self-Calibration, or press
RUN
key to start the self-calibration.
Self Calibration
Note: Warm up the oscilloscope at least 30 minutes
before performing self-calibration. Do not have
anything connected to any of the inputs. Doing
so will create errors and instrument may fail to
calibrate properly.