Menu Operation
82
Press the
More 1/2
softkey to display the
UTILITY
menu
page 2/2.
Softkey
Options
Description
Service
----
Select
Service
menu
Pass/Fail
----
Select
PASS/FAIL
menu
Self-Cal
RUN/STOP
Start self-calibration
AUTO
Exit self-calibration
Fast-Cal
ON
Fast calibrate the
vertical position
OFF
Turn off fast calibration
More 2/2
----
Select menu page 1/2
Self-Calibration
If you want to maximize measurement accuracy, you can
perform a self-calibration.
Self-calibration uses the internally generated signals to
optimize circuits that affect channel scale, offset and trigger
parameters for all the divisions over the full range.
Disconnect all inputs and allow the oscilloscope to warm up
at least 30 minutes before performing self-calibration.