Quick Start Guide
212
2. Press the
MENU
key and select
Output Type
to be
User ARB.
3. Then, select
More 1/2
softkey and select
Capture/Storage
softkey.
4. Press the
Source
softkey and select A+B.
Note: This selection will change depending on the
Operate parameter set under MATH menu.
5. Select
Data Type
as Period Data if you want to store
a complete period of the CH1+CH2 waveform, or
select Screen Data to capture only what’s displayed
on the screen.
6. Then press
Internal Storage
and change
Storage
Pos
to the memory location to store the waveform to.
7. Press
Save
and the CH1+CH2 waveform will be
saved into internal arbitrary waveform memory, which
can be recalled and output.
Note: The captured Math waveform will be limited to
600 points, and some scaling is done internally.
See the “Capture/Storage” section for details.