Quick Start Guide
200
6.7 Making Cursor Measurements
You can use the cursors to quickly make time and voltage
measurements on a waveform. You can use the cursors to
measure the amplitude and frequency of a FFT waveform.
You can also use the cursors to measure the phase
difference between two signals with the same frequency
when X-Y horizontal mode is selected.
Measure the time and voltage on normal waveform
Perform the following steps to take time and frequency
measurements.
Press the
CURSOR
key to display the
CURSOR
menu.
Press
Mode
softkey to select the
Manual
mode.
Press
Type
softkey to select the
Time
type.
Press
X1--/
X2—
softkey or press the entry knob to
select X1 cursor.
Rotate the entry knob
to move the X1 cursor.
Press
X1--/
X2—
softkey or press the entry knob to
select X2 cursor.
Rotate the entry knob
to move the X2 cursor.
ΔX and 1/ΔX are displayed in the softkey area. ΔX is
the time difference between X1 and X2;
1/ΔX is the
frequency between X1 and X2.
Summary of Contents for 2540B
Page 1: ...Model 2540B 2542B 2540B GEN 2542B GEN Digital Storage Oscilloscopes USER MANUAL ...
Page 2: ...2 ...
Page 30: ......
Page 45: ...Basic Operation 45 Vertical Invert off Vertical Invert on ...
Page 64: ...Basic Operation 64 Roll Horizontal Mode ...
Page 138: ...Menu Operation 138 3 6 DISPLAY Menu Display Menu key ...
Page 146: ......
Page 184: ......
Page 213: ...Remote Control 213 7 REMOTE CONTROL Comsoft Software Web Browser GUI Graphic user Interface ...
Page 215: ...Troubleshooting Guide 215 8 TROUBLESHOOTING GUIDE System Messages General Problems ...
Page 222: ......
Page 248: ......
Page 249: ......