Measure the phase difference between two signals of the
same frequency under X-Y display mode.
Connect a sine wave signal to CH1 and a sine wave
signal of the same frequency but out of phase to CH2.
Press horizontal
MENU
key to display the
Horizontal
menu.
Press
X-Y
softkey to select
X-Y
display mode
Center the signal on the display with the vertical control
knob of each channel.
Use the vertical scale control knob of each channel to
expand the signal for convenient view.
Press the
CURSOR
key to display the
CURSOR
menu.
Press
Mode
softkey to select the
Manual
mode.
Press
Source
softkey to select
CH2
.
Press
Type
softkey to select the
Voltage
type.
Press
Y1--/
Y2—
softkey or press the entry knob to
select Y1 cursor.
Rotate the entry knob
to move the Y1 cursor to the
top of the signal.
Press
Y1--/
Y2—
softkey or press the entry knob to
select Y2 cursor.
Rotate the entry knob
to move the Y2 cursor to the
bottom of the signal.
ΔY displayed in the softkey area is the voltage difference
D (or 2B) between Y1 and Y2.
Summary of Contents for 2540B
Page 1: ...Model 2540B 2542B 2540B GEN 2542B GEN Digital Storage Oscilloscopes USER MANUAL ...
Page 2: ...2 ...
Page 30: ......
Page 45: ...Basic Operation 45 Vertical Invert off Vertical Invert on ...
Page 64: ...Basic Operation 64 Roll Horizontal Mode ...
Page 138: ...Menu Operation 138 3 6 DISPLAY Menu Display Menu key ...
Page 146: ......
Page 184: ......
Page 213: ...Remote Control 213 7 REMOTE CONTROL Comsoft Software Web Browser GUI Graphic user Interface ...
Page 215: ...Troubleshooting Guide 215 8 TROUBLESHOOTING GUIDE System Messages General Problems ...
Page 222: ......
Page 248: ......
Page 249: ......