Performance Verification Procedure
243
Time Scale Accuracy
Equipment needed: Fluke calibrator (preferred) or signal generator
1. Connect your calibrator’s output to CH1 of the oscilloscope
by 50ohm feed thru connector. If recommended calibrator
is not available, connect alternative equipment as follows:
2. On the oscilloscope, the time base is switched to the sweep
speed under test.
3. Set up a 10 MHz sine wave output from calibrator.
4. Press the Auto button on oscilloscope to get a stable
waveform.
5. Press the Measure button, soft key
Time
, and then soft key
Frequency
.
6. Press the Acquire button and soft key
Mode
to set
Average
mode.
7. Adjust average to 8. It should read 10 MHz.
8.
Switch sweep speed to 100 ms (or 200 ms) to have a
stable frequency reading. This reading should be less than
Oscilloscope
Calibrator or
Signal