Page 5
5
Explanation of functions and settings
5.1
General description of the remote control
5.2
Function of the «Reglobeam» remote control
5.6
Configuration with keys
5.3
Layout of the «Reglobeam» remote control
5.4
Configuration mode
5.5
Establishing the connection
The Reglobeam remote control allows you to program Herkules 2 easily
and conveniently from the ground. Data transfer between Reglobeam and
Herkules 2 functions in both directions, i.e. to and from the detector, and
is guaranteed by an infrared interface. The Reglobeam reads back the
adjusted values immediately after programming and displays them for
control purposes. This ensures safe and correct programming.
Please refer to the first part (4.6) for an overview of individual parameters
and the enhanced level settings.
Note:
Programming must be carried out within 30 s. If this time elapses,
it is necessary to activate programming mode again (press key G
The «Reglobeam» remote control functions using a combination of func-
tions and number keys. Take care to operate it correctly as described in
chapter 4.2.
Flashing keys on the Reglobeam indicate that the data has not been fully
transmitted. Avoid exposing the infrared interface to direct sunlight or
other light sources.
The connection between the Reglobeam and Herkules 2 can only be
established when the detector is in configuration mode. Configuration
mode is activated when the detector is switched on. For safety reasons,
this mode is automatically deactivated 30 minutes after the last setting
has been made on the detector. Configuration mode can be exited at any
time by pressing keys A+4.
Configuration mode can be activated in three different ways:
a)
By pressing any key X or Y key on the detector
Changing the function
Press key X. The value increases by 1 for every key press (function no.).
Once the last function has been reached, the program jumps back to the
first function. The green LED then indicates the number of the activated
function.
Changing the level
Press key Y. The value increases by 1 for every key press (level). Once the
last level has been reached, the program jumps back to the first level.
Exiting programming mode
Exit programming mode by pressing both keys X and Y briefly at the same
time.
X
Y
Key
Key
X
Y
Key
Key
b)
By restarting the detector (disconnecting the supply voltage)
c)
Access via remote control with access code
Press start key G on the Reglobeam
➔
If the connection has been established successfully, G and one of the
keys 1 to 7 light up (address of the detector)
➔
If G flashes, the connection could not be established
➔
Hold the remote control closer to the detector and point directly at it
➔
Check the batteries of the Reglobeam.
➔
Replace the batteries if nothing lights up.
➔
Configuration mode is not activated if G and one of the keys 1 to 7
light up although no further settings are possible.
Primary mode:
Press keys X and Y simultaneously and hold for 2 s
➔
Green LED lights up, then release
Secondary mode:
Press keys X and Y simultaneously and hold for 4 s
➔
Red LED lights up, then release
Reset:
Press keys X and Y simultaneously and hold for 8 s
➔
Both LED lights up, then release
Mode
Parameter (key X)
Function no. Level key Y
Primary
Enhanced level
1
1–8
Mounting height
2
1–7
Relay parameter setting
3
1–8
Field size
4
1–9
Sekundär
Relay hold interval
1
1–5
Direction recognition
2
1–3
Crossing traffic masking
3
1–9
Wide field
4
1–2
SMD level
5
1–9
SMD field size
6
1–9
Digital filter function
7
1–2
Numerical keys
to
1
9
Battery
compartment
Infrared interface
Function keys
to
F
A
Locking
screw
Start
G
Front side
Rear side
Procedure
Press keys X and Y at the
same time to access the
required mode.