6.1 Battery Indicator
The battery indicator is located in the lower RIGHT corner of the LCD. The
indicator is comprised of 4 bars representing the amount of power left in the
battery. When 4 bars appear, the battery is fully charged.
If LOW BATTERY appears on the LCD display mid treatment, the System
Indicator Light on the top of the neurostimulator will flash YELLOW, an
audible beep will occur for 5 seconds and then once per minute until the end
of the treatment. The battery should have enough charge in it to allow the
patient to complete the remainder of the treatment.
At the end of the treatment, when the battery
is almost depleted, the neurostimulator will
indicate LOW BATTERY on the LCD display
along with the Maximum Intensity reached
during the treatment. LOW BATTERY means
that the stimulator cannot complete the next
treatment with the remaining charge left on the battery and accordingly, it
will not allow a treatment to start until the battery has been recharged.
6.2 Charging the Battery
First, unplug the leadwire cable from the neurostimulator, revealing the
opening for the plug on the AC Charger. Make sure BiowavePRO is turned
off. Plug the AC Charger into the charging port on the neurostimulator. Plug
one end of the power cord into the AC Charger and the other end into an
electrical wall outlet to begin charging the battery. After about one minute,
the bars inside the large picture of the battery on the LCD display will begin
animating and the System Indicator Light above the display will begin
flashing
GREEN indicating that the battery is now being charged.
Biowave Li-ion batteries take approximately 4.5 hours to fully recharge from
empty.
6. Battery Indicator, Charging
and Replacing the Battery