Biowave Corporation
103
9. Technical Specifications
and Classifications
The BiowavePRO
®
Neurostimulator is intended for use in the electromagnetic
environment specified below. The customer or the user of the BiowavePRO
Neurostimulator should assure that it is used in such an environment.
10 V
10 V/m
3 V (RMS) 150
kHz to 80 MHz
3 V/m 80 MHz
to 2.5 GHz
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Immunity test
IEC 60601
test level
Compliance
level
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Electromagnetic
environment – guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the BiowavePRO Neurostimulator,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter as provided below.
Recommended separation
distance:
d = 0.35
√
P
d = 0.35
√
P 80 MHz to 800 MHz
d = 0.7
√
P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance
in meters (m). Field strengths from
fixed RF transmitters, as determined
by an electromagnetic site survey
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol: