Error Codes
|
179
Synergy HT Operator's Manual
Code
Description and Probable Causes:
0600
Filter gain out of range
Fail if PMT saturation is not met within 3 ms.
This error indicates that during fluorescence autocalibration, the gain was increased to an unsafe
level, and the PMT analog PCB did not reach saturation.
Probable Causes:
•
The autocalibration jig is incorrectly placed in the carrier. The mirrors are not facing in the
correct direction.
•
The EM autocalibration jig is not connected to the PMT analog PCB.
0601 - 0606
Time-Resolved function reference channel gain out of range for the selected
wavelength
Fail if reference signal > 40000 and gain = 1.
This error indicates that the absorbance reference channel gain for a specific wavelength during
Time-resolved readings is out of the range necessary to ensure the lambda performance to
specifications. The second to last number is the lambda table position number.
Probable Causes:
•
The absorbance reference channel PCB is defective.
•
Order-sorting filter wheel is jammed not aligning the correct bandpass filter with the light
path, or the through hole is lined up, allowing white light to pass.
•
Monochromator is defective.
•
Flash lamp, alignment, or Flash lamp power supply is defective.
•
Lamp is too bright.
0610 - 0660
Time-Resolved function reference channel gain out of range for the selected
wavelength
Fail if Reference signal > 40000 and Gain = 1.
Note:
The order of the last two digits is <Filter> <Channel>. This is not consistent with other
error codes.
This error indicates that the absorbance reference channel gain for a specific wavelength during
Time-Resolved readings is out of the range necessary to ensure the lambda performance to
specifications. The second to last number is the lambda table position number.
Probable Causes:
•
The absorbance reference channel PCB is defective.
•
Order-sorting filter wheel is jammed, not aligning the correct bandpass filter with the light
path, or the through hole is allowing white light to pass.
•
Monochromator is defective.
•
Flash lamp, alignment, or Flash lamp power supply is defective.
•
Lamp is too bright.
0611 - 0616
Measurement channel gain out of range for the selected wavelength
Fail if Air reading > 60000 with a Gain = 1.
This error indicates that during a filter calibration or testing the reader prior to a read, one of the
filters saturated the measurement channel. The last number is the lambda table position
number.
Probable Causes:
•
The monochromator mirror/grating is damaged.
•
Order-sorting filter wheel is jammed, not aligning the correct bandpass filter with the light
path; the filter is degraded and not passing enough light energy, or filter is blocking the
light.
•
Absorbance analog PCB has intermittently failed.
•
Missed flashes or an erratic flash lamp.
•
Voltage to the lamp has decreased due to failure of the lamp power supply, or the motor
power supply PCB sent the wrong voltage request to the lamp power supply.
Summary of Contents for Synergy HT
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Page 2: ......
Page 22: ...xx Preface BioTek Instruments ...
Page 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Page 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Page 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Page 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Page 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...