Routine Procedure for Models With Injectors
|
159
Synergy HT Operator's Manual
Rinse the Fluid Lines
Perform this procedure only if decontamination was performed using sodium hypochlorite.
1.
Place a beaker containing at least 30 ml of deionized water on the dispense
module.
2.
Place the SYRINGE 1 or 2
inlet tube
in the beaker.
3.
If you have not already done so, place the
outlet tubes
in an empty
beaker.
4.
From the
Reader Control
dialog, select
Dispenser 1
or
2
, set the
Volume
to
5000 µl
, and keep the default dispense
Rate
.
5.
Run
five
prime cycles, for a total of 25000 l.
6.
Pause for
10 minutes
and then run
one
prime cycle with 5000 µl. This
delay will allow any residual sodium hypochlorite to diffuse into the
solution and be flushed out with the next prime.
7.
Empty
the beaker containing the outlet tubes.
8.
Wipe all surfaces with deionized water.
9.
Discard the used gloves and cloths using a Biohazard trash bag and an
approved Biohazard container.
Clean the Internal Tubing and Injector Heads
Turn to
Chapter 5, Preventive Maintenance
and perform the following
procedures to access, remove, and clean the internal tubing and injector heads:
•
Required Materials
•
Removing the Reader’s Shroud
•
Removing the Internal Tubes and Injector Heads
•
Cleaning the Internal Tubes and Injector Heads
When finished, replace the internal components and the reader’s shroud.
Summary of Contents for Synergy HT
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Page 2: ......
Page 22: ...xx Preface BioTek Instruments ...
Page 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Page 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Page 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Page 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Page 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...