Gen5™ Software
|
51
Synergy HT Operator's Manual
Purge Utility
Gen5™ provides a special utility to purge fluid from the dispense tubing and
syringe by pumping the fluid in reverse, back into the supply bottle. To purge the
dispense module:
1.
In Gen5, select
System|Reader Control|Synergy(Com<#>)
and click
the
Dispenser
tab.
2.
Select the
Dispenser
number (1 or 2) associated with the supply bottle.
3.
Enter the desired purge
Volume
in L.
4.
Select a prime
Rate
in L/second.
5.
Click
Purge
to start the process.
Syringe Maintenance Position
Gen5 provides access to special syringe setup functions for maintenance and
calibration purposes. If a syringe needs to be installed or replaced, it must first be
moved to its “Maintenance Position.” To do this using Gen5:
1.
In Gen5, select
System|Reader Control|Synergy(Com<#>)
and click
the
Dispenser
tab.
2.
Select the appropriate
Dispenser
number (1 or 2) associated with the
syringe.
3.
Click
Maintenance.
The syringe plunger will move to its furthest-from-
home position. The syringe can then be disconnected form the drive bracket
and unscrewed from the valve.
4.
See “Install Dispense Module Components” in
Chapter 2, Installation
for information on installing/removing the syringes.
L
Important!
Do not change the syringe positions or
calibrate the dispensers unless instructed to do so as part
of installation, upgrade, or maintenance.
Summary of Contents for Synergy HT
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Page 2: ......
Page 22: ...xx Preface BioTek Instruments ...
Page 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Page 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Page 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Page 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Page 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...