Cleaning the Internal Components
|
143
Synergy HT Operator's Manual
Once the shroud has been removed and the internal tubes and injector heads have
been removed and cleaned (see page 140), follow these instructions to remove a
few more components and then clean the optical probes:
1.
Disconnect the heater and thermistor wires. To do this, depress the small
tab (pictured below) and separate the connectors.
2.
Remove the thumbscrew located in the left rear of the instrument and set it
aside. This exposes the ground wire.
Depress tab to separate
the connectors