Syringe Maintenance Position
KC4 provides access to special syringe setup functions for maintenance and
calibration purposes. If a syringe needs to be installed or replaced, it must first be
moved to its “Maintenance Position.” To do this using KC4:
1.
In KC4, select
System|Readers,
click the
Configuration
button, and then
click the appropriate
Dispenser
tab (#1 or #2).
2.
Click
Move Syringe to maintenance position.
The syringe plunger will move to its furthest-from-home position. The
syringe can then be disconnected form the drive bracket and unscrewed
from the valve. See “Install Dispense Module Components” in
Chapter 2,
Installation
for information on installing/removing the syringes.
3.
Click
OK
to close the dialog.
L
Important!
Do not change the syringe positions or
calibrate the dispensers unless instructed to do so as part
of installation, upgrade, or maintenance.