| 11
BioTek Instruments, Inc.
3.
Connect a bottle containing deionized water and run DAY_RINSE three
times to flush the system.
Clean Components
Periodic Maintenance
Periodic maintenance involves cleaning the washer components on a
regular basis to keep the washer running efficiently and in compliance with
instrument specifications. The recommended
frequency for cleaning
washer components
is
at least monthly
. Risk and performance factors
associated with your assays may require performing some or all of the
procedures more frequently.
Warning! Internal Voltage.
Always turn off and disconnect
the instrument from its power supply for all cleaning
operations.
Important!
Do not immerse the instrument, spray it with
liquid, or use a “wet” cloth. Do not allow the cleaning solution
to run into the interior of the instrument. If this happens,
contact the BioTek Service Department.
Do not soak the touchscreen—
this will cause damage.
Moisten a clean cloth with deionized or distilled water and
wipe the touchscreen. Dry immediately with a clean, dry
cloth.
Do not apply lubricants
to manifold gasket, manifold
channel end seals, fittings, tubing connections, or any
surface that is part of the fluid path. The use of any lubricant
on the fluid handling components will interfere with the
aspirating and dispensing performance, and may cause
irreparable damage to these components.
Maintenance Tools and Supplies
•
70% isopropyl alcohol
•
Deionized or distilled water
•
Dish soap or other mild cleaner
•
Tools for cleaning dispense and aspirate tubes on
page 14
•
Phillips head screwdriver
•
Lint-free disposable towels