18 |
50™ TS Washer
Self-Test to check the manifold, microplate carrier, vacuum pump and syringe
pump positioning and operation.
When the self test has completed successfully, the washer is ready for use.
If the test passes
, the Main Menu is displayed.
If the test fails
, the washer beeps repeatedly and displays an error code. If
this happens, note the error code and then press
Stop
on the keypad to stop
the beeping. Look up
Error Codes on page 1
to determine its cause. If the
problem is something you can fix, turn off the washer, fix the problem, and
then turn the washer back on. Otherwise, contact BioTek’s Technical
Assistance Center.
The
Instrument Testing section
provides recommended procedures to
perform after the instrument is installed and set up as described in this
chapter, and
before
the instrument is used in a laboratory environment.
The successful completion of the Installation Qualification confirms that the
washer and its components have been supplied as ordered and ensures that
they are assembled and configured properly for your lab environment. The
successful completion of the Operational Qualification confirms that the
washer is operating according to specification.
Note:
An
instrument qualification package
(PN 155) for the 50 TS
is available for purchase from BioTek. The package contains thorough
procedures for performing Installation Qualification, Operational
Qualification and Performance Qualification (IQ/OQ/PQ) and
preventive maintenance (PM). Extensive Checklists and Logbooks are
included for recording results. Contact your local dealer for more
information.
Important!
Review
Optimize Performance recommendations in the
operator's manual
, which describes necessary steps to perform before
running the washer.