BioTek Instruments, Inc.
Specify the Interval and AutoPrime Parameters
AutoPrime only runs when the Main Menu (Home) is displayed and the
instrument has been idle for a specified interval.
To set
AutoPrime
:
Touchscreen:
LHC:
1.
Select
Instrument
at the main menu and
select
AutoPrime
.
1.
Select
Tools>Instrument
Utilities> AutoPrime
tab.
2.
“Enable” AutoPrime and
specify the parameters:
•
the idle-time interval that
will trigger priming; up to
24 hours.
•
Set the
Submerge
Duration
, if desired.
•
Flow rate, volume, and
buffer valve, if
applicable.
3.
LHC users: click
Send
to
transfer the settings to the
instrument.