| 15
BioTek Instruments, Inc.
Disconnect the waste tube from the Carrier port when you are not
using the vacuum filtration carrier. An internal valve preserves the
vacuum required to perform standard aspiration.
Installation Final Check
1.
Verify that the tubing was not crimped during installation.
2.
Make sure the fluid supply and the waste output tubes are attached to
the appropriate manifold ports. Lift the manifold mounting bracket up
or back to make sure the tubing to the manifold is not being pulled too
tightly.
3.
Vacuum Filtration models: make sure there is sufficient room behind
the washer for the waste tubing attached to the special carrier to move
during operation.
Attach the Mist Shield
The mist shield is also called the aerosol cover.