25
System Settings
»
Line Flush:
If enabled, the system flushes the column inlet
line at the end of the run using the run’s weakest solvent.
Select whether to
Collect
the flush (using the same collection
criteria as during the gradient run) or send it
To Waste
.
Note:
If the automatic line flush is disabled (
Off
), ensure to
flush the column flow path manually before each run.
»
Detector Flush:
If enabled, the system flushes the UV detector
at the end of the run using the run’s strongest solvent.
Select whether to
Collect
the flush (using the same collection
criteria as during the gradient run) or send it
To Waste
.
Note:
We recommend that the detector flush is enabled.
A contaminated detector flow cell has decreased
transmissivity, which causes increased noise levels,
decreased response, and difficulties performing UV Zero.
Optional Detectors
»
Biotage
®
Solvent Detector:
Enable this option when the
optional solvent detector(s) from Biotage is connected
to the system. Leak detection using up to three solvent
detectors is supported on all systems that show three
detectors (
) in the
About
view when the this
option
is enabled. For more information, see “Leak Detection
Using Biotage
Network (Optional)
N
See “Configure Network Share and E-mail Functionality” on
page 26.
Licenses
To enable
λ
-all detection mode and baseline correction, you
need a Spektra software license. To purchase a Spektra
software license, please contact your local representative.
To install a license:
1. Create a directory called \biotage\selekt\ on a USB memory
device and save the license file to this location.
2. Connect the memory device to a USB port on the left side of
the system.
3. Press
Install
in the
Licenses
field.
4. When a message saying that the license was successfully
installed appears, remove the USB memory device.