21
4. Alarm threshold
The alarm threshold is set to 38°C. This cannot be changed. If the
temperature is over 38°C, a beep sounds during the temperature
measurement.
5. Changing the batteries
When the LCD screen displays
, the battery is used.Operation:
Open the cover and replace the used batteries taking care to observe
polarity. A polarity error may cause damage and void the guarantee.
Do not use batteries other than the batteries mentioned, do not
recharge non-rechargeable batteries, or throw them into a fire.
Remove the batteries when not using the device for a long period. Use
disposable batteries.
VIII. PRECISION
ACCORDING
TO ASTM
STANDARD
1965-1998
(2009)
34°C - 35.9°C =
± 0.3°C
93.2°F - 96.6°F =
± 0.3°F
36°C - 39°C = ± 0.2°C
96.8°F - 102.2°F =
± 0.2°F
39°C - 42.5°C = ±
0.3°C
102.2°F - 108.5°F =
± 0.3°F
The
THERMOFLASH® LX-26E
can take temperature measurements
below 32°C / 89.6°F or above 42.9°C / 109.22°F. Outside these
temperature ranges, precision is not guaranteed.
Longevity of use
The
ThermoFlash® Premium
is guaranteed for about 40,000 readings.
IX. ACCESSORIES SUPPLIED
User manual, LR6 (AA)*2PCS.
X. TROUBLESHOOTING
If you experience any of these problems using your
THERMOFLASH®
Premium
, please refer to this troubleshooting guide to resolve the
problem. If the problem persists, please contact our customer service at
+33 892 350 334 (0.40€ per minute).
- The display indicates temperatures above 95°:
the temperature is
indicated in Fahrenheit. Change the unit of measure to Celsius.
Summary of Contents for Exacto ThermoFlash Premium
Page 2: ...Premium INSTANT MEASUREMENT LESS THAN 1 SECOND MESURE INSTANTAN E MOINS DE 1 SECONDE...
Page 6: ...5 MANUEL D UTILISATION FR Premium MESURE INSTANTAN E MOINS DE 1 SECONDE...
Page 14: ...15 Premium INSTANT MEASUREMENT LESS THAN 1 SECOND USER MANUAL EN...
Page 22: ...117 E L E C T R O M A G N E T I C C O M PAT I B I L I T Y D ATA...
Page 29: ...30...
Page 30: ...125 WA R R A N T Y 31...