Bionova® IC10/20FR

 

Hygiene monitoring systems

Readout results for ISO 15883-5 Threshold Mode

The ISO 15883-5:2021 standard indicates the use of two 

different  protein  quantity  thresholds  to  evaluate  cleaning 
efficacy: An Action threshold, and an Alert threshold. ISO 15883-

5 based thresholds are expressed as ratios in µg/cm

units that 

relates protein quantity found in a sample (µg) and the swabbing 

surface (cm

2

) on which the sample was taken.  According to the 

standard, each threshold has different implications:

Alert threshold: 

The alert levels are the target levels of 

cleaning  efficacy  that  the  cleaning  process  should  achieve.  

Recommended threshold value 3µg/cm

2

. Lower values may be 

recommended  for particular applications.

Action threshold: 

The action levels are the maximum criteria 

for  acceptable  cleaning  efficacy  during  testing  of  a  sample. 

Recommended threshold value 6.4µg/cm

2

. Lower values may 

be recommended  for particular applications.

Values between the two thresholds should be investigated, 

but are considered to pass cleaning requirements. 

Important:

 Refer to the ISO 15883-5:2021 standard for 

detailed information on how to implement the standard to 
your processes. 

If ISO 15883-5 Threshold Mode is selected as the readout 
threshold mode on your Auto-reader, each time a readout is 

completed the protein quantity will be informed by the Auto-

reader through the Quantitative Protein Level LED Indicators 
and through a printed ticket:

Approximate amounts of

BSA expressed in μg

1 2 3 4 5

1 µg

2 µg

3 µg

4 µg

5 µg

Also, as a quick visual cue of the protein quantity found on the 

sample, the device will turn on the Red LED Indicator if the 

protein quantity is above 1 µg, or the Green Indicator if the 
protein quantity is below 1 µg.

Important:

 Please keep in mind that according to ISO 15833-

5:2021  standard  the  protein  quantity  found  in  a  sample 

(in µg) needs to be related to the swabbed surface area to 
obtain a readout result (in µg/cm

2

) that can be comparable 

with the selected Alarm and Alert thresholds (also in µg/cm

2

).

Once  the  protein  quantity  in  the  sample  has  been  obtained 

by the Auto-reader, and the swabbed surface information has 
been obtained with the Bionova

®

 Antinova APP. The information 

is finally linked together by the Bionova

®

 Cloud Environment to 

inform the readout result. Please refer to 

ISO 15883-5:2021 on 

Bionova

®

 Cloud Environment

 section for a step by step guide to 

upload and link data on Bionova Cloud Environment.

Once the data has been linked, the readout result  will be 
shown on the 

Hygiene Monitoring > Finished

 tab of Bionova

®

 

Cloud Web Interface. Depending on the threshold values 
selected, the informed readout result can be as follows: 

If the protein quantity found on the swabbed surface (µg/cm

2

) is 

below the Alert Threshold. The readout result will be informed 
as 

NEGATIVE

.

If the protein quantity found on the swabbed surface (µg/cm

2

is above the Alert Threshold but below the Action Threshold. 
The readout result will be informed as 

ALERT

.

If  the  protein  quantity  found  on  the  swabbed  surface  (µg/

cm

2

) is above the Action Threshold. The readout result will be 

informed as 

ACTION

.

Take action immediately whenever an 

ACTION

 Readout 

Result  is  informed  as  this  indicate  that  the  protein  quantity 
found does not comply with the user defined quality standards 

set for the cleaning process.

Disposal

Discard the PRO1 indicator according to your country’s sanitary 
regulations. It is not possible to use or incubate a PRO1 indicator 
more than once. Please refer to the PRO1 instructions for use.

Audible alarm

An audible alarm will sound every time a positive result is 
detected in your Auto-reader. The alarm allows the user 
to immediately detect a positive result without the need to 
visually control the device. The alarm can be canceled by 
pressing the 

 button .

Cancel a reading

The user can cancel a reading by pressing the

button 

and the 

button for three seconds. The device will cancel 

the reading and the auto-reader will print out a ticket 

confirming the cancellation.

Summary of Contents for IC10/20FR

Page 1: ...WARNING this product contains dry natural rubber Rev 31 February 2022 Bionova IC10 20FR Auto reader for Rapid Super Rapid and Ultra Rapid Biological Indicators and Hygiene Monitoring systems...

Page 2: ......

Page 3: ...results 9 Audible alarm 9 Temperature monitoring 9 Thermal paper specifications 9 Replacement of the paper roll 10 Set up mode 10 Date format and time zone selection mode 11 Time setting mode 12 Print...

Page 4: ...of the sterilization processes should be incubated at 60 C Simultaneous incubation of SCBIs for monitoring EO processes along with SCBIs for monitoring other sterilization processes is not possible Pl...

Page 5: ...to high luminous intensity lamps Do not place the incubator near devices that emit strong electromagnetic fields Do not use the incubator on leaned surfaces or on surfaces that are subject to shocks...

Page 6: ...he user Operating conditions Power supply specifications Input parameters Operating conditions Units Voltage range 100 240 Volts Frequency 50 60 Hertz Current 0 2 Amperes Output parameters Values Unit...

Page 7: ...selection mode Press again to select the desired temperature After 4 seconds the temperature modification will be accepted automatically 5 Select the appropriate incubation time for the SCBI you are p...

Page 8: ...efore the selected time limit as part of its normal operation Please refer to the Product compatibility and incubation settings section for more information When a negative result is detected in a rea...

Page 9: ...se The positive control and the exposed SCBI should belong to the same batch Disposal Discard the SCBIs according to your country s sanitary regulations Positive SCBIs can be sterilized before discard...

Page 10: ...here are three types of date format DD MM YY MM DD YY and YY MM DD Press the or button to move back and forth between each of the 3 available options After 2 seconds a ticket with the selected format...

Page 11: ...C and start up the program Select the Bionova IC10 20FR Auto reader and press the Clock Sync button the auto reader will be synchronized with the PC s clock Synchronization can only be done as long as...

Page 12: ...button To decrease Instructions for use ten minutes value press button Note Possible values for ten minutes are 0 1 2 3 4 and 5 Minutes setting Unit minutes hh mM Press button to turn on the blue ligh...

Page 13: ...dications detailed on Threshold Mode selection on Bionova Cloud Environment section All devices are manufactured with the following settings Default Threshold Mode HTM 01 01 Default Protein Threshold...

Page 14: ...alues selected on your Auto reader Please refer to the Interpretation of results of PRO1 indicators section for more information After a Readout result is informed remove the indicator from the incuba...

Page 15: ...otein level upper limit of acceptable protein contamination after processing is 5 g BSA equivalent per instrument side indicating as well that lower levels may be recommended for particular applicatio...

Page 16: ...ce area to obtain a readout result in g cm2 that can be comparable with the selected Alarm and Alert thresholds also in g cm2 Once the protein quantity in the sample has been obtained by the Auto read...

Page 17: ...pplication and the Bionova Cloud Agent which is a Microsoft Windows App that acts as an interface between the Terragene compatible devices and the web based application Bionova Cloud Traceability Soft...

Page 18: ...and swabbing surface data in cm2 informed by the Surface Eye App throughout the Bionova Cloud Environment allows the creation of ratios of protein quantities and swabbing surface area in g cm2 that ca...

Page 19: ...on The APP will upload all the surface information to your Bionova Cloud Environment account Bionova Cloud Environment 8 The surface acquisition details can be edited from the Apps home menu swiping t...

Page 20: ...nd the protein quantities that correspond to the alert and action thresholds for the defined swabbing area Bionova Cloud Environment Cleaning and maintenance Cleaning and decontamination of external s...

Page 21: ...The printer cover is not tightly closed Printer without paper Paper roll is not placed correctly Bionova Cloud Agent is running on the PC The Auto reader is turned off The Auto reader is not connected...

Page 22: ...The warranty period for electronic products marketed by Terragene will be 1 one year from the date of first use of the product and may never be extended beyond the product s useful life 5 five years f...

Page 23: ......

Reviews: