Bionova® IC10/20FR 

7

 

Safety information

 Instructions for use 

Start-up

1|

Place the Bionova

®

 IC10/20FR Auto-reader Incubator 

on  a  firm  surface,  free  from  vibrations,  away  from  direct 

sunlight, currents of hot or cold air, chemical and corrosive 

or  flammable  substances.  Do  not  place  the  Incubator  in  a 

way that disconnection of the power supply's AC plug from 

mains could be difficult. Leave a space of at least 10 cm from 

the Incubator to the closest wall. Do not move the Incubator 
periodically or during its use. Connect the Incubator to a 
secure and stable electrical mains outlet socket.

Do  not  wet  or  heat  the  device.  If  liquid  is  spilled  on  the 

Incubator, disconnect it and follow the instructions on the 

Cleaning and maintenance

 section.

2|

Power on your Bionova

®

 IC10/20FR by connecting the 

power supply's AC plug to mains and then connect the plug 
at the other end of power supply to the rear of the Incubator. 
The last incubation program used (time and temperature 
combination) will be selected by default.

Before power on, verify that all incubation positions are empty.

If the Printer's paper indicator light starts blinking after 
powering on the Incubator, check that the printer's door is 
tightly closed and that the printer has paper, if not follow the 
instructions on the 

Replacement of the paper roll

 section.

3|

Set the incubator internal clock to correspond to your local 

time. See configuration methods in the 

Time setting mode 

section.

All devices are manufactured with the following settings:

 Time zone: UTC +0:00
 Printing language: English

If you want to change the time zone, or the printing language 
refer to 

Modification of time zone

, or 

Modification of printing 

language

 on the 

Set-up mode

 section.

4|

 Select the appropriate incubation temperature for the SCBI 

you are planing to use, according the SCBI’s Instructions for 
use. Press and hold the temperature selection button for 3 
seconds to enter the temperature selection mode. Press 
again to select the desired temperature. After 4 seconds the 

temperature modification will be accepted automatically. 

5| 

Select the appropriate incubation time for the SCBI you are 

planning to use by pressing the left incubation time selection 
button (for selecting the incubation time for incubations positions 
1to 6), or the right incubation time selection button ((for selecting 
the incubation time for incubations positions 7 to 12).
Please refer to the 

Product Compatibility and Incubation 

programs

 section to know more about the temperature and time 

combinations (incubation programs) available for your device.

5|

Wait until the temperature stability indicator light stops blinking 

for the incubator to reach a steady working temperature.

!

Do not place an indicator in an incubation position before the 
incubation temperature is stable.

6|

Once the temperature stability indicator light remains 

stable, a one-time automated test to check each position 
internal status will be carried out (auto-test). Once the test is 
complete, the position status indicator light will turn green to 
indicate that the position can be used to perform readings, or 
it will turn red to indicate that an error has occurred.
When the incubator detects an error in a position, the position 
will be disabled to guarantee the reliability of the results and it 
cannot be used to perform any further readings.
To indicate that a position has been disabled, the position 

Temperature             Time 

       Incubation

selection                   selection 

       program

60 ºC 

                 20 min 

(Ultra)              

20 min at 60 ºC

 

                 30 min 

       30 min at 60 °C

 

                 1 hs 

       1 hour at 60 °C

 

                 2 hs                             2 hours at 60 °C

 

                 3 hs 

       3 hours at 60 °C

 

                 PRO (7 min) 

       PRO at 60 °C

37 ºC 

                 4 hs 

       4 hours at 37 °C

To select the appropriate incubation program for the Indicator 
you are planning to use, follow each Indicator’s Instructions 
for use.

Select the appropriate incubation program before the 
incubation of any Indicator.

Bionova® IC10/20FR Auto-reader 

compatible indicators

We develop new Indicators for extending the features of 
our incubators regularly. Please visit 

www.terragene.com/

IC1020FR

 

to  find  out  all  compatible  indicators  for  your 

device.

Summary of Contents for IC10/20FR

Page 1: ...WARNING this product contains dry natural rubber Rev 31 February 2022 Bionova IC10 20FR Auto reader for Rapid Super Rapid and Ultra Rapid Biological Indicators and Hygiene Monitoring systems...

Page 2: ......

Page 3: ...results 9 Audible alarm 9 Temperature monitoring 9 Thermal paper specifications 9 Replacement of the paper roll 10 Set up mode 10 Date format and time zone selection mode 11 Time setting mode 12 Print...

Page 4: ...of the sterilization processes should be incubated at 60 C Simultaneous incubation of SCBIs for monitoring EO processes along with SCBIs for monitoring other sterilization processes is not possible Pl...

Page 5: ...to high luminous intensity lamps Do not place the incubator near devices that emit strong electromagnetic fields Do not use the incubator on leaned surfaces or on surfaces that are subject to shocks...

Page 6: ...he user Operating conditions Power supply specifications Input parameters Operating conditions Units Voltage range 100 240 Volts Frequency 50 60 Hertz Current 0 2 Amperes Output parameters Values Unit...

Page 7: ...selection mode Press again to select the desired temperature After 4 seconds the temperature modification will be accepted automatically 5 Select the appropriate incubation time for the SCBI you are p...

Page 8: ...efore the selected time limit as part of its normal operation Please refer to the Product compatibility and incubation settings section for more information When a negative result is detected in a rea...

Page 9: ...se The positive control and the exposed SCBI should belong to the same batch Disposal Discard the SCBIs according to your country s sanitary regulations Positive SCBIs can be sterilized before discard...

Page 10: ...here are three types of date format DD MM YY MM DD YY and YY MM DD Press the or button to move back and forth between each of the 3 available options After 2 seconds a ticket with the selected format...

Page 11: ...C and start up the program Select the Bionova IC10 20FR Auto reader and press the Clock Sync button the auto reader will be synchronized with the PC s clock Synchronization can only be done as long as...

Page 12: ...button To decrease Instructions for use ten minutes value press button Note Possible values for ten minutes are 0 1 2 3 4 and 5 Minutes setting Unit minutes hh mM Press button to turn on the blue ligh...

Page 13: ...dications detailed on Threshold Mode selection on Bionova Cloud Environment section All devices are manufactured with the following settings Default Threshold Mode HTM 01 01 Default Protein Threshold...

Page 14: ...alues selected on your Auto reader Please refer to the Interpretation of results of PRO1 indicators section for more information After a Readout result is informed remove the indicator from the incuba...

Page 15: ...otein level upper limit of acceptable protein contamination after processing is 5 g BSA equivalent per instrument side indicating as well that lower levels may be recommended for particular applicatio...

Page 16: ...ce area to obtain a readout result in g cm2 that can be comparable with the selected Alarm and Alert thresholds also in g cm2 Once the protein quantity in the sample has been obtained by the Auto read...

Page 17: ...pplication and the Bionova Cloud Agent which is a Microsoft Windows App that acts as an interface between the Terragene compatible devices and the web based application Bionova Cloud Traceability Soft...

Page 18: ...and swabbing surface data in cm2 informed by the Surface Eye App throughout the Bionova Cloud Environment allows the creation of ratios of protein quantities and swabbing surface area in g cm2 that ca...

Page 19: ...on The APP will upload all the surface information to your Bionova Cloud Environment account Bionova Cloud Environment 8 The surface acquisition details can be edited from the Apps home menu swiping t...

Page 20: ...nd the protein quantities that correspond to the alert and action thresholds for the defined swabbing area Bionova Cloud Environment Cleaning and maintenance Cleaning and decontamination of external s...

Page 21: ...The printer cover is not tightly closed Printer without paper Paper roll is not placed correctly Bionova Cloud Agent is running on the PC The Auto reader is turned off The Auto reader is not connected...

Page 22: ...The warranty period for electronic products marketed by Terragene will be 1 one year from the date of first use of the product and may never be extended beyond the product s useful life 5 five years f...