176
BM3Vet Pro
User’s Manual
Guidance and manufacturer’s declaration
- electromagnetic immunity
The BM3Vet Pro system is intended for use in the electromagnetic environment specified below.
The customer or the user of the BM3 Vet Pro system should assure that it is used in such an
environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment -guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80MH
z
3 Vrms
150 kHz to 80 MH
z
BM3 Vet Pro system must be used only in
a shielded location with a minimum RF sh
ielding effectiveness and, for each cable t
hat enters the shielded location with a mi
nimum RF shielding effectiveness and, for
each cable that enters the shielded locati
on
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.5
GHz
3 V/m
80.0 MHz to 2.5 G
Hz
Field strengths outside the shielded locati
on from fixed RF transmitters, as determin
ed by an electromagnetic site survey, sho
uld be less than 3V/m.
a
Interference may occur in the vicinity of e
quipment marked with the following sym
bol:
Note 1)
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Note 2)
It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to assure that they meet the minimum specification.
Summary of Contents for BM3Vet Pro
Page 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Page 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Page 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...