65
BM3Vet Pro
User’s Manual
cancel the protection provided by the isolated input. In particular, there must be no contact
of the neutral electrode and ground.
DEFIBRILLATION
—
Do not come into contact with animals during defibrillation. Otherwise
serious injury or death could result.
To avoid the risk of serious electrical burn, shock, or other injury during defibrillation, all
persons must keep clear of the bed and must not touch the animal or any equipment
connected to the animal.
After defibrillation, the screen display recovers within 10seconds if the correct electrodes are
used and applied in accordance with the manufactur
er’s
instructions.
Patient cables can be damaged when connected to an animal during defibrillation. Check
cables for functionality before using them again.
The peak of the synchronized defibrillator discharge should be delivered within 60ms of the
peak of the R wave. The signal at the ECG output on the veterinary patient monitor s is
delayed by a maximum of 30ms.
If the ECG waveform on the screen is too unstable to synchronize with the animal
’
s heart
beat because of the following reason, remove the cause of an alarm, message, or unstable
ECG, and then use a stable ECG lead for synchronization.
✓
ECG electrode is detached or broken. Lead wire is detached or broken.
✓
Lead wire moves. AC interference, EMG noise or noise from ESU is superimposed.
✓
Connection cable is broken or has a short circuit. Connector has poor contact.
INTERFACING OTHER EQUIPMENT
—
Devices may only be interconnected with each other
or to parts of the system when it has been determined by qualified biomedical engineering
personnel that there is no danger to the animal, the operator, or the environment as a
result. In those instances where there is any element of doubt concerning the safety of
connected devices, the user must contact the manufacturers concerned (or other informed
experts) for proper use. In all cases, safe and proper operation should be verified with the
applicable
Manufacturer’s instructions for use and system
standards IEC 60601-1-1/EN 60601-1-1 must
be complied with.
Electro surgery Unit
Summary of Contents for BM3Vet Pro
Page 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Page 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Page 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...