181
BM3Vet Pro
User’s Manual
Sidestream CO2 (Option)
Measurement range
0 to 150 mmHg, 0 to 19%
Accuracy
0-40mmHg
±
2 mmHg
41-70mmHg
±
5% of reading
71-100mmHg
±
8% of reading
101-150mmHg
±
10% of reading
Respiration rate
2 to 150 breath per minute
Respiration accuracy
±
1 breath per minute
Mainstream CO2 (Option)
Measurement range
0 to 150 mmHg, 0 to 19%
Accuracy
0-40mmHg
±
2 mmHg
41-70mmHg
±
5% of reading
71-100mmHg
±
8% of reading
101-150mmHg
±
10% of reading
Respiration rate
0 to 150 breath per minute
Respiration accuracy
±
1 breath per minute
Anesthetic GAS/O2 - Phasein (Option)
Method
Infra-red absorption characteristic
Gas
CO2, O2, N2O, Des, Iso, Enf, Hal, Sev
Warm-up time
Mainstream (IRMA AX+)
Iso accuracy mode: 45s
Full accuracy mode: 60s
Side stream (ISA OR+/ AX+): <20s
Sample flow rate
(for ISA OR+/AX+)
50 ± 10 ml/min
Measuring Range
CO2: 0 ~ 15 %
N2O: 0 ~ 100%
Hal/Iso/Enf: 0 ~ 8%
Sev: 0~10%
Des: 0 ~ 22%
O2: 0 ~ 100% (ISA OR+/AX+)
Respiratory Rate
0 ~ 150bpm
±
1bpm