configured on the network switch. Bionet equipment must be configured to make the network
settings compatible with the specifications of the operating organization.
This device exchanges data with other medical devices over a LAN network. The network supports
the following transports and protocols:
⚫
TCP / IP
⚫
Broadcast
VLAN Network
If data is exchanged within a single network, an independent VLAN network for the clinical
information system must be established. At least one of the following independent VLAN
networks must be established.
⚫
Network for medical devices in hospital
⚫
Network for portable veterinary patient monitors
If you use an inappropriate network
If your network does not meet the requirements, the following dangerous situations can occur.
If the distributed alarm system is not safe:
⚫
The alarm will not be delivered.
⚫
The alarm or data is delayed.
⚫
An error alarm appears
If the network connection is interrupted:
⚫
The alarm will not be delivered.
⚫
Reactivates with the alarm off or the alarm sound off
If you do not have firewall and antivirus software:
⚫
Your data is not protected.
⚫
The device settings are changed.
⚫
The device raises an error alarm or does not generate an alarm.
⚫
Data is sent incomplete, to the wrong device, or not at all.
⚫
Animal patient data is blocked, falsified, or corrupted.
⚫
The time stamp of the data is inaccurate
Overloading this unit due to very high network loading (e.g. denial of service attacks) can cause
Summary of Contents for BM3Vet Pro
Page 22: ...22 BM3Vet Pro User s Manual Left Side 1 Printer 1...
Page 45: ...45 BM3Vet Pro User s Manual Display Mode Wave and Numeric Mode Numeric Mode...
Page 68: ...68 BM3Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 146: ...146 BM3Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 204: ...204 BM3Vet Pro User s Manual less than minus number percent plus or minus...