BM1VOM-1.24
8
.EtCO2 95
BM1VET Operation Manual
The airway adapters for sidestream intubated applications
3473ADU-00
Airway Adapter
Kit w/
Dehumidification
Tubing
Weight: 4.5 grams
Deadspace
– adds approximately 7
cc of deadspace
Intended for use when
monitoring animals with ET
Tube sizes >4.0 mm
3473INF-00
Airway Adapter
Kit w/
Dehumidification
Tubing
Weight: 5.8 grams
Deadspace
– adds approximately 1
cc of deadspace
Intended for use when
monitoring animals with ET
Tube sizes <=4.0 mm
-
Connecting the LoFlo Sample Kit
1. The sample cell of the sampling kit must be inserted into the sample cell receptacle of the LoFlo
CO
2
Module as shown in Figure 1. A “click” will be heard when the sample cell is properly inserted.
2. Inserting the sample cell into the receptacle automatically starts the sampling pump.
Removal of the sample cell turns the sample pump off.
3. To remove the sampling kit sample cell from the sample cell receptacle, press down on the
locking tab and pull the sample cell from the sample cell receptacle.