Cleaning Applied Parts
Cables and Leadwires
CAUTION
Do not use acetone or ketone solvents for cleaning; do not use an autoclave or steam cleaner.
Cables and leadwires can be cleaned with a warm, damp cloth and mild soap, or isopropyl alcohol
wipes. For more intensive disinfecting (near sterile) Ethylene Oxide (ETO) is acceptable but will
reduce the useful lifetime of the cable or leadwire.
CAUTION
The decision to sterilize must be made per your institution’s requirements with an awareness of
the effect on the integrity of the cable or lead wire.
Note
The Equipment needs a
safety inspection once a year. Please refer to user’s guide or service
manual for the exam specifications.
Please check carefully both frame and sensor, after cleaning the Equipment, Do not use equipment
that is worn out or damaged.
At least once a month, clean and wipe off the frame by using a soft cloth after wetting it with water
and alcohol. Do not use lacquer, thinner, ethylene, or oxidizer which may lead to damage to the
equipment.
Make sure both cables and accessories are free of dust or contaminants, and wipe them off with soft
cloth wetted with warm water (40°), and at least once a week, clean them by using clinical alcohol.
Do not submerge the accessories under any liquid or detergent. Also, make sure that no liquid
penetrates into the Equipment or probes.
Caution
Do not dispose of single use probe in any hazard place, Always think about environmental
contamination.