BM1VOM-1.24
5.SpO2 67
BM1VET Operation Manual
⚫
Position of SpO
2
Probe
BM1Vet offers multisite SpO
2
probe and transflectance SpO
2
probe as standard accessories. When
multisite SpO2 probe is used, the preferred sensor application site for canine, feline and equine
animals is on the tongue,
with the sensor’s optical components positioned on the center of the
tongue. Alternatively, the sensor and clip may be applied to the animal’s lip, toe, ear, prepuce, or
vulva. When transflectance SpO2 probe is used, the preferred sensor application site for canine,
feline and equine animals is in the rectal part or under the tail.
WARNING
Use only Bionet's reusable multisite SpO
2
probe or transflectance SpO
2
probe. Use of other
oxygen transducers may cause improper performance.
Note
The signal input is a high-insulation port and it is defibrillator proof ( )
The insulated input ensures animal safety and protects the device during defibrillation and
electrosurgery.