BM1VOM-1.24
1.BASIC
15
BM1VET Operation Manual
CAUTIONS
Intended Use
This device is designed to be used for monitoring the biological vital signs of dogs and cats. Main
functions of the product include displaying information such as SpO
2
, NIBP, carbon dioxide (CO
2
)
and temperature on its LCD screen and monitoring parameter, and alarming.
Application Environment
This device is for use by trained veterinary personnel in veterinary centers. The device is restricted
to be used on one patient at a time.
Operator Requirement
Only veterinary personnel who have read the Operator’s Manual should use this monitor.
Instruction for Use
For continued safe use of this equipment, it is necessary that the instructions are followed. However,
instructions listed in this manual in no way supersede established medical practices concerning
animal care.
Loss of Veterinary Vital Signs
Should the monitor at any time temporarily lose vital signs from animal , the potential exists that
active monitoring is not being done. Close animal observation or alternate monitoring devices should
be used until monitor function is restored.
If the monitor does not automatically resume operation within 60 seconds, power cycle the monitor
using the power on/off switch. Once monitoring is restored, you should verify correct monitoring state
and alarm function.
Maintenance
Regular preventive maintenance should be carried out annually (Technical inspections). You are
responsible for any requirements specific to your country.
MPSO
The use of a multiple portable socket outlet (MPSO) for a system will result in an enclosure leakage
current equal to the sum of all individual earth leakage currents of the system if there is an
interruption of the MPSO protective earth conductor. Do not use an additional extension cable with
the MPSO as it will increase the chance of the single protective earth conductor interruption.
Negligence
BIONET does not assume responsibility for damage to the equipment caused by improperly vented
cabinets, improper or faulty power, or insufficient wall strength to support equipment mounted on
such walls.