Operation and Care Manual
Scenario
A: The unit will not operate. No fans blowing, with a blank screen.
1. Check all power connections to the InnerVue
™
II unit. Check the power cable
connection to the wall outlet and the rear of the InnerVue
™
II unit.
See § 3.2.2 Setup for Each Procedure
2. Ensure that the voltage-selector switch is in the appropriate position.
See § 3.2.2 Setup for Each Procedure
3. Check fuses.
See § 4.5 Fuse Replacement
4. Check the building breaker controlling the outlet you are using. You may need
to consult the proper building administration to ensure power to your outlet.
5. Should these measures fail, please contact Biomet Sports Medicine for technical
support and/or service.
B: The unit’s fans are blowing but the monitor is blank.
1. Ensure the camera cable connector is properly seated in the receptacle on the
front of the unit.
See § 3.2.2 Setup for Each Procedure
2. Check for light output. If there is none, replace the lamp.
See § 4.3 Lamp Replacement
3. Contact InnerVue
™
II technical support and/or service. Make a note of the camera
you are using. If you have multiple camera attachments, the support person
may ask you to try another camera. It may be necessary to send the camera you
are using for service along with the InnerVue
™
II unit.
C. The image is blurry.
1. If you are using a camera coupler, rotate the focus ring to bring the image into
focus.
2. The scope, coupler or camera optics may be dirty. Please refer to the manuals
that came with your camera and scope for proper care and cleaning.
D. There is no light or light output is low.
1. If light output seems low, check the connection between the light output on
the camera and the scope for particles and foreign objects.
2. If the output is still low or there is no light, your lamp may be bad. Please
contact your service representative.
See § 4.3 Lamp Replacement
3. Ensure the scope and handpiece are properly aligned.
4. If your lamp is good or is newly replaced, and light output is now still low,
the light fibers in the camera cable may be damaged. Please contact Biomet
Sports Medicine for service.